Where the equation comes from
Dilution means adding solvent to a stock solution so the amount of solute stays constant while the total volume increases. The amount of solute before equals the amount after: C1 × V1 = C2 × V2.
C1 is the stock concentration, V1 is the volume of stock you take. C2 is the concentration you want after dilution, V2 is the final total volume of the diluted solution. Rearranged for the unknown stock volume: V1 = (C2 × V2) / C1.
Worked example: molar concentrations
You have a 10 mM stock and need 100 mL of 1 mM. C1 = 10 mM, C2 = 1 mM, V2 = 100 mL. V1 = (1 × 100) / 10 = 10 mL of stock. Diluent = 100 − 10 = 90 mL. Dilution factor = 10 / 1 = 10 (a 10-fold dilution).
The BiomeBench Dilution Calculator returns these three values (stock volume, diluent volume, dilution factor) for any valid inputs.
Worked example: percent concentrations
You have 70 % ethanol and need 500 mL of 40 % ethanol. C1 = 70 %, C2 = 40 %, V2 = 500 mL. V1 = (40 × 500) / 70 ≈ 286 mL of 70 % stock. Diluent = 500 − 286 = 214 mL. Dilution factor = 70 / 40 = 1.75.
Percent (v/v or w/v) and fold (×) units are treated as consistent ratios and are not interconverted by the calculator.
Worked example: fold concentrations
A 100× stock needs to become 1× working solution. C1 = 100×, C2 = 1×, V2 = 1 L. V1 = (1 × 1000 mL) / 100 = 10 mL. Diluent = 1000 − 10 = 990 mL. Dilution factor = 100 / 1 = 100.
The dilution factor
The dilution factor is C1 / C2, or equivalently V2 / V1. It tells you how many times the working solution is more dilute than the stock.
A 10-fold dilution has a factor of 10: you take 1 part stock and make the total volume up to 10 parts (so 1 part stock + 9 parts diluent). Do not confuse the factor with the 1:9 mixing ratio — the factor counts the final total volume.
In a serial dilution each step multiplies the previous concentration by the inverse of the factor. After n steps the concentration is the stock concentration divided by factorⁿ.
Unit conversion table
| Quantity | 1 unit equals |
|---|---|
| 1 M | 1,000 mM = 1,000,000 µM = 1,000,000,000 nM |
| 1 g/L | 1 mg/mL = 1 µg/µL |
| 1 mg/mL | 1,000 µg/mL = 1,000,000 ng/mL |
| 1% w/v | 10 mg/mL = 10 g/L |
| 1% v/v | 1 mL per 100 mL = 10 µL per mL |
| 10× | 10-fold dilution |
| 100× | 100-fold dilution |
Common mistakes and how to avoid them
Prefix mismatch: entering C1 in mM and C2 in µM makes V1 off by 1,000×. Always use the same unit for both concentrations.
Confusing V2 with diluent volume: V2 is the final solution volume. Diluent = V2 − V1, not V2.
Using the equation for concentration: C1V1 = C2V2 only works when C2 < C1 (dilution). To concentrate a solution you remove solvent, which this equation does not model.
Not mixing before measuring V1: an unmixed stock gives a wrong concentration in the aliquot.
When a serial dilution is better
When the dilution factor is larger than about 100, the required stock volume becomes too small to pipette accurately (below about 2 µL). A serial dilution uses easy, repeatable volumes at each step instead. The Serial Dilution Planner applies C1V1 = C2V2 repeatedly across a series and reports every tube.