The equation
C1V1 = C2V2 relates a starting solution to a diluted one. C1 is the concentration of the stock, V1 is the volume you take from it, C2 is the concentration you want after dilution, and V2 is the final total volume.
Rearranged for the volume to move: V1 = (C2 × V2) / C1. This is what a dilution calculator solves when you ask for a target concentration in a final volume.
How the dilution factor enters
The dilution factor is C1 / C2, or equivalently V2 / V1. A ten-fold dilution has a factor of 10: you move one part of sample into nine parts of diluent, so V2 is ten times V1. A two-fold dilution has a factor of 2: one part sample into one part diluent.
In a serial dilution each step multiplies the previous concentration by the inverse of the factor. After n steps the concentration is C_stock divided by factor^n.
The cumulative factor after n identical steps, from a 1 mg/mL stock:
| Steps (n) | Ten-fold series: factor | Ten-fold: concentration | Two-fold series: factor | Two-fold: concentration |
|---|---|---|---|---|
| 1 | 10 | 0.1 mg/mL | 2 | 0.5 mg/mL |
| 2 | 100 | 0.01 mg/mL | 4 | 0.25 mg/mL |
| 3 | 1,000 | 0.001 mg/mL | 8 | 0.125 mg/mL |
| 4 | 10,000 | 0.0001 mg/mL | 16 | 0.0625 mg/mL |
| 5 | 100,000 | 0.00001 mg/mL | 32 | 0.03125 mg/mL |
| 6 | 1,000,000 | 0.000001 mg/mL | 64 | 0.015625 mg/mL |
Dilution factor from the volumes you mix
You can also read the dilution factor straight from the volumes you mix: factor = final volume / sample volume. Move 100 µL of sample into a final volume of 1,000 µL and the factor is 1,000 / 100 = 10.
That equals C1 / C2; both describe the same step, so use whichever form you have values for.
Common mistakes and how to avoid them
Most serial dilution errors come from the setup, not from the math. A small mix-up in one tube carries into every tube after it.
Confusing the dilution factor with the dilution ratio is the most common slip. Some notes write 1:10 to mean one part sample plus nine parts diluent; others write it to mean one part sample in ten parts total. Write the actual transfer volume and the final volume instead of a ratio alone.
Another error is using the whole mixed volume for V1 in C1V1 = C2V2. V1 is only the volume of sample you transfer, not the total after mixing.
Worked example: ten-fold dilution
You want 10 mL of a 10^-3 dilution from a 1 × 10^6 cells per mL stock. C1 = 1 × 10^6, C2 = 1 × 10^3, V2 = 10 mL. V1 = (1 × 10^3 × 10) / (1 × 10^6) = 0.01 mL, or 10 µL. So move 10 µL of stock into 9.99 mL of diluent.
Built as a serial tube series instead: tube 1 gets 100 µL of stock into 900 µL of diluent (10^-1). Tube 2 gets 100 µL from tube 1 into 900 µL (10^-2). Tube 3 gets 100 µL from tube 2 into 900 µL (10^-3).
Worked example: two-fold dilution
You want a two-fold series from a stock of 1 mg per mL, ending at 0.0625 mg per mL. Each step halves the concentration: 1, 0.5, 0.25, 0.125, 0.0625. Four dilution steps, or 2^4 = 16-fold overall.
For each tube, move an equal volume of stock or previous tube into an equal volume of diluent. For example, 500 µL into 500 µL gives a two-fold dilution.
Reading the log scale
Serial dilutions are exponential, not linear. On a log scale each tube sits at an even spacing: log10(10^-1) = -1, log10(10^-2) = -2, and so on. A straight line on a semi-log plot means the dilutions were accurate; a curve means a step was off.
When you plate for CFU, plot the count against the dilution factor. The linear part of that curve gives the concentration in the original sample.
When to use a serial dilution
A serial dilution earns its place when you need a wide range of lower concentrations from one starting sample, or when a direct dilution would be too large to measure well.
Turning a strong stock into a very weak standard may need a 1,000-fold or 10,000-fold reduction. A single step at that scale needs a tiny sample volume, which raises pipetting error. A stepwise dilution uses easy, repeatable volumes instead.
A direct dilution is enough when the target is close to the starting concentration and the required volume is easy to measure accurately.
- A test range across several concentrations.
- Better accuracy than one very large dilution step.
- Repeatable standards for assays or calibration curves.
- Clear concentration tracking across many samples.