What a master mix is

A PCR master mix is one tube holding every reagent shared by all reactions: buffer, dNTPs, MgCl₂, polymerase, primers, and water. You aliquot the mix into tubes and add template to each tube separately. Every tube gets the same mix, so the only difference between tubes is the template.

Preparing reactions one reagent at a time is slow and error-prone, especially across 24, 48, or 96 tubes. A master mix makes every tube identical except for its template, and the overage you prepare keeps the last tube from running short.

Worked example: the 25 µL recipe

This is the standard Taq recipe the calculator scales. It is illustrative: verify stocks and your polymerase’s own protocol before use. Research use only.

Each concentration-based component follows V = final × reaction volume ÷ stock (C1V1 = C2V2). Polymerase is the exception: volume = units per reaction ÷ stock (U/µL). Water = reaction volume − template − all additives. Additives excluding template total 7.25 µL; with 1.00 µL template and 16.75 µL water the reaction is 25.00 µL.

Do not pipette the per-reaction volumes directly. Taq at 0.25 µL and dNTPs at 0.50 µL per tube are at or below accurate pipetting range. That is what the master mix is for.

  • 10× PCR buffer: 10× stock, 1× final, 2.50 µL per reaction
  • dNTP mix: 10 mM each stock, 0.20 mM each final, 0.50 µL per reaction
  • MgCl₂: 25 mM stock, 1.50 mM final, 1.50 µL per reaction
  • Forward primer: 10 µM stock, 0.50 µM final, 1.25 µL per reaction
  • Reverse primer: 10 µM stock, 0.50 µM final, 1.25 µL per reaction
  • Taq polymerase: 5 U/µL stock, 1.25 U per reaction, 0.25 µL per reaction
  • Template DNA: added separately, 1.00 µL per reaction
  • Nuclease-free water: to volume, 16.75 µL per reaction
  • Total: 25.00 µL per reaction
Worked example: the 25 µL recipe Worked example: the 25 µL recipe — 25.00 µL per reaction: water 16.75, buffer 2.50, MgCl₂: 1.50, primers 1.25 + 1.25, dNTPs 0.50, Taq 0.25, template 1.00 µL 25.00 µL 16.75 µL 2.50 µL 0.50 µL 1.50 µL 1.25 µL 1.25 µL 0.25 µL 1.00 µL × 10 (+10%) = 264.00 µL

PCR Master Mix Calculator

Scaled totals: 10, 24, and 96 reactions

The table scales the recipe with 10% overage: multiply each per-reaction volume by the number of reactions and by 1.1. Water and every component scale by the same factor. Template is not scaled into the mix; it is added per tube afterwards.

Volumes below are for 10, 24, and 96 reactions, in that order. Per-reaction master mix excluding template is 24.00 µL, so the totals are 264.00 µL, 633.60 µL, and 2,534.40 µL (2.53 mL).

96 reactions at 10% overage does not fit a 1.5 mL tube. Use a 5 mL tube, or split the mix, or drop to 5% overage if reagent cost matters: 96 reactions at 5% is 2,419.2 µL, which still needs a large tube. A fixed 10% works for up to 50 reactions; above that, think about the tube first.

ComponentPer reaction10 reactions (+10%)24 reactions (+10%)96 reactions (+10%)
10× buffer2.50 µL27.50 µL66.00 µL264.00 µL
dNTPs0.50 µL5.50 µL13.20 µL52.80 µL
MgCl₂1.50 µL16.50 µL39.60 µL158.40 µL
Forward primer1.25 µL13.75 µL33.00 µL132.00 µL
Reverse primer1.25 µL13.75 µL33.00 µL132.00 µL
Taq polymerase0.25 µL2.75 µL6.60 µL26.40 µL
Water16.75 µL184.25 µL442.20 µL1,768.80 µL
Total (excl. template)24.00 µL264.00 µL633.60 µL2,534.40 µL

Order of addition

Add water first, then buffer, dNTPs, MgCl₂, primers, and polymerase last. Mix gently and spin down briefly. Dispense the mix, then add template to each tube separately. Keep polymerase on ice or in a cold block and return it promptly. Do not vortex polymerase vigorously.

Check whether your buffer already contains MgCl₂. Many commercial buffers do. Do not add the separate MgCl₂ from the table on top of a buffer that already contains it without checking the datasheet. The calculator treats extra MgCl₂ as optional: set its final concentration to 0 to omit it.

This recipe is for separate components, not a 2× premix. If you use a 2× master mix, buffer, dNTPs, and polymerase are already in it. Set each of their final concentrations to 0 in the calculator to omit them, and do not double-add them. The calculator removes an omitted component from the mix and rebalances the water.

Annealing temperature

The BiomeBench Oligo Melting Temp calculator implements the Wallace rule (13 nt or fewer) and the basic GC formula (14 to 70 nt) only. It reports Tm, length and GC content. It does not implement nearest-neighbour thermodynamics, has no salt-concentration input, and gives no self-dimer warning. Do not treat its Tm as a primer-design Tm.

Use the primer supplier’s nearest-neighbour Tm as the starting point for the annealing temperature (Ta). For Taq, a common starting rule is 5 °C below the lowest primer Tm, while NEB’s guideline for most polymerases other than Q5 and Phusion is 0 to 2 °C below the lowest primer Tm. Either way, verify with a gradient, and do not cite NEB for the 5 °C rule.

For Q5 and other high-fidelity polymerases, NEB puts the starting point 0 to 3 °C above the lowest primer Tm. Run a gradient around that starting point for a new primer pair, and keep the pair’s Tm values within about 3 °C of each other. If they differ by more, redesign the primers rather than forcing the annealing temperature.

Troubleshooting

Each row is a starting diagnosis, not a diagnosis. Work through the likely causes in order, change one thing at a time, and keep the no-template control clean.

  • No band. Likely causes: template missing or too little, primer or polymerase omitted, annealing too high, or the wrong primer pair. Try: check the template on a gel or by A260, re-check the mix against the calculator, or drop Ta by 2 to 3 °C or run a gradient.
  • Smear or multiple bands. Likely causes: annealing too low, too much template, too many cycles, or primer-dimer and non-specific priming. Try: raise Ta, reduce template and cycles, or check primer design. The Tm calculator does not check dimers; use a dedicated primer tool.
  • Primer-dimer band below about 100 bp. Likely causes: primer concentration too high, primers self-complementary, or annealing too low. Try: reduce primer to 0.2 to 0.3 µM final, redesign the primers, or raise Ta.
  • Weak or faint band. Likely causes: too little template or polymerase, degraded dNTPs, or annealing too high. Try: increase template within the water balance, use fresh dNTPs, or lower Ta by 2 to 3 °C.
  • Band in the no-template control. Likely cause: contamination. Try: fresh water and reagents, a clean workspace, and new primers. Do not trust the run until the control is clean.

From master mix to plate

Dispense the mix into the plate or tubes, then add template to each well or tube separately. The mix volume per well comes from the calculator; in this recipe the template is the same 1 µL per well.

Plan the plate layout in the 96-Well Experiment Planner before you pipette. It checks volume and spatial consistency across the plate, so a misloaded well shows up before the run, not after.

Where the recipe appears on this site

The BiomeBench PCR Master Mix Calculator implements this recipe. Enter the per-reaction volumes of every component, the number of reactions, and the excess fraction, and it returns the total volume of each component and the final master-mix volume, with warnings when template or primer amounts cross well-known bench thresholds. It does not check primer Tm matching or dimer formation, and it does not model gradient optimisation.

Use the Oligo Melting Temp calculator for primer Tm and the Nucleic Acid Concentration calculator to check template amount by A260 before the run.