UV-Vis absorbance (A260)
A spectrophotometer measures absorbance at 260 nm. For double-stranded DNA, an A260 of 1.0 in a 1 cm cuvette corresponds to about 50 micrograms per mL. The calculation is simple and needs no dye, but absorbance at 260 nm also picks up RNA, free nucleotides, and other contaminants, so it overestimates DNA in impure samples.
The A260/A280 ratio checks purity: about 1.8 is clean double-stranded DNA. A lower value suggests protein or phenol contamination; a higher value can mean RNA contamination.
Nanodrop-style microvolume spectrophotometry
A microvolume spectrophotometer such as Nanodrop needs only 1 to 2 microlitres and reports concentration plus the A260/A280 and A260/A230 ratios. It is fast and convenient, but like all absorbance methods it cannot tell DNA from RNA, and it is unreliable below about 2 to 5 nanograms per microlitre.
Use it for a quick purity check on a relatively clean sample. Do not rely on it for low-concentration or heavily contaminated DNA.
Qubit fluorometry
A Qubit fluorometer uses a dye that binds selectively to double-stranded DNA. It does not respond strongly to RNA or free nucleotides, so it reads DNA accurately even in contaminated samples. The working range is roughly 10 picograms per µL to 100 nanograms per µL for the high-sensitivity assay.
The trade-off is that it needs a dye, standards, and a dedicated instrument. It takes longer than a spectrophotometer reading and costs more per sample.
Use Qubit when you need an accurate DNA concentration for downstream work such as sequencing library preparation, PCR setup, or transfection.
PicoGreen fluorescence
PicoGreen is a fluorescent dye with high selectivity for double-stranded DNA. It is more sensitive than UV absorbance and works in plate-reader format, so it suits high-throughput screening of many samples.
Like Qubit it needs a dye and a fluorescence reader, and it reads double-stranded DNA specifically. Single-stranded DNA and RNA give much weaker signals.
Comparison at a glance
| Method | What it measures | Sample volume | Selective for dsDNA? | Purity info | Working range |
|---|---|---|---|---|---|
| UV-Vis (A260) | All nucleic acids + contaminants | Cuvette | No | A260/A280 ratio (1.8 DNA, 2.0 RNA) | — |
| NanoDrop | All nucleic acids + contaminants | 1–2 µL | No | A260/A280 ratio | Unreliable below ~2–5 ng/µL |
| Qubit | dsDNA only | Dye + standards, µL volumes | Yes | None | 10 pg/µL – 100 ng/µL (high-sensitivity assay) |
| PicoGreen | dsDNA only | Plate-reader format | Yes | None | Low, plate-reader scale |
When to use each method
Choose absorbance (UV-Vis or Nanodrop) for a fast, reagent-free check when the sample is reasonably clean and you also want the purity ratios.
Choose Qubit or PicoGreen when accuracy matters more than speed, especially for low-concentration DNA, contaminated samples, or anything going into a sensitive downstream assay.
Never mix methods across a single experiment: an absorbance value and a fluorescence value for the same tube can differ by a factor of two or more, and both can be right within their own limits.
Link to the quantification calculator
The BiomeBench Nucleic Acid Quantification calculator converts A260 absorbance into DNA or RNA concentration in micrograms per mL, applies dilution factors, reports the A260/A280 purity ratio, and calculates total yield. It implements the A260 method described here.