Skip to main content

Nucleic Acid Quantification

Convert a spectrophotometer A260 reading into DNA or RNA concentration using the standard absorbance factors (50 µg/mL for dsDNA, 40 µg/mL for RNA, 33 µg/mL for ssDNA). Enter the dilution factor used for the reading, and optionally an A280 to compute the A260/A280 purity ratio. Copy-number mode converts a mass concentration and template length into copies per µL and plans the 10-fold qPCR standard series.

Local context

Nucleic-acid concentration is usually read from absorbance at 260 nm. The conversion from absorbance to concentration is a single multiplication by a published factor — but the factor differs between DNA and RNA, and between single- and double-stranded molecules: dsDNA = 50 µg/mL per AU, ssDNA = 33 µg/mL per AU, RNA = 40 µg/mL per AU (at 1 cm pathlength). The A260/A280 ratio indicates purity (1.8–2.0 for DNA, ~2.0 for RNA); A260/A230 indicates organic/contaminant carryover.

This calculator applies the right factor for the molecule you measured and converts between absorbance, mass concentration (µg/mL, ng/µL), and total yield (µg, ng) when you supply the sample volume. It accepts 1 cm or 0.1 cm (NanoDrop) pathlength and auto-scales the factor.

When NOT to use: for protein concentration from A280, use the Protein Concentration Calculator which uses the protein's specific extinction coefficient (or sequence-derived estimate). For oligo Tm, use the Oligo Melting Temp Calculator. For converting between mass and molar concentration of a known sequence, use the Molarity Calculator with the calculated MW.

Key formula: C (µg/mL) = A260 × factor × dilution. Factor: dsDNA 50, ssDNA 33, RNA 40. Reference: Sambrook & Russell (2001) Molecular Cloning; Thermo Fisher NanoDrop technical guide.

Calculator

Inputs

A260 quantification converts an absorbance reading into concentration. Copy number converts a mass concentration into copies per µL for a qPCR standard curve.

Blank-corrected absorbance reading at 260 nm.

Sets the conversion factor: 50 µg/mL (dsDNA), 33 (ssDNA), or 40 (RNA) per 1.0 A260.

Total dilution before reading (e.g. 2 µL sample in 100 µL → 50).

Used only for the A260/A280 purity ratio.

Total sample volume, for computing total yield in µg.

Enter a reading and calculate to see the concentration

Related guides

About this calculation

What this is

Nucleic acids absorb at 260 nm, and the absorbance of a 1 mg/mL solution in a 1 cm path length is a published constant (50 for dsDNA, 40 for RNA, 33 for ssDNA). Measuring A260 in a diluted aliquot and multiplying by the dilution factor and the right constant yields the stock concentration.

Why it is used

Every downstream application (PCR, sequencing, transfection, in vitro transcription) needs an accurate nucleic acid concentration. A260 is the most accessible bench measurement, but the constant and dilution must be applied correctly.

Why this calculator exists

Mixing up the absorbance factor (50 vs 40 vs 33) or forgetting the dilution factor are the two classic bench errors. The calculator makes the constant explicit, multiplies by your dilution, and reports a purity ratio when you also provide an A280.

Purity ratios at a glance

Read the ratios the calculator reports against this table. A260/A280 checks protein contamination; A260/A230 checks organic carryover such as phenol, guanidine salts, or carbohydrates.

RatioPure dsDNAPure RNALow value suggests
A260/A2801.7–2.0 (about 1.8)1.9–2.1 (about 2.0)Protein or phenol (at or below 1.6)
A260/A2302.0–2.22.0–2.2Salt, carbohydrate, or phenol carryover

Common mistakes

  • Confusing solvent volume with final solution volume. Prepare to the final volume. The dilution factor applies to the final solution, not the solvent alone.
  • Skipping the mix step before reading. Mix the sample. An un-mixed sample gives an uneven reading.
  • Reading a sample that contains bubbles or particulates. Clear bubbles and particulates first. They scatter light and read artificially high.
  • Comparing readings taken at different path lengths without correcting. Correct for path length. The conversion factors assume a 1 cm path.
  • Reading a sample above the linear range. Dilute the sample and re-read instead of trusting a saturated value. Absorbances above about 1.5 to 2.0 deviate from linearity.
  • Using the wrong A260 factor for the analyte. Use 50 µg/mL per A260 unit for double-stranded DNA, 40 for RNA, and 33 for single-stranded DNA.

Key assumptions

  • The spectrophotometer was blanked correctly with the same buffer the sample was diluted into.
  • The sample is pure enough at A260 for Beer–Lambert to apply (A260 < ~1.5).
  • The path length is 1 cm (or your instrument has been corrected for a different path length).

How to use it

Enter the blank-corrected A260 reading, then choose the analyte: double-stranded DNA, single-stranded DNA, or RNA. This sets the conversion factor (50, 33, or 40 µg/mL per A260 unit).

Include the total dilution you applied before reading. If you mixed 2 µL of sample into 100 µL, the dilution factor is 50, not 98.

Enter A280 as well to get the A260/A280 purity ratio. Use it as a coarse check, not a verdict: a ratio outside the typical range means the sample needs cleaning, not that the numbers are wrong.

Supply the final volume to get the total yield in µg. This is the number to record before you dilute or aliquot a precious sample.

If the reading is above the linear range of your instrument, dilute the sample and re-read rather than trusting a saturated value. Multiply the result back by the new dilution.

Check the displayed unit carefully. A concentration in µg/mL and a yield in µg are easy to confuse, and the difference matters when you plan a downstream reaction.

Switch the mode to Copy number to convert a mass concentration into copies per µL. Enter the concentration in ng/µL and the template length; the calculator applies the molar mass per base (660 dsDNA, 330 ssDNA, 340 RNA) and Avogadro’s number.

Copy-number mode also plans the 10-fold qPCR standard series from your stock down to 100 copies/µL, with the pipetting volumes for each standard.

Limitations

  • A260 cannot distinguish DNA from RNA: use a fluorometric assay for that.
  • Contaminants such as phenol or carbohydrates inflate A260, producing overestimated concentrations.
  • Very low concentrations (A260 < 0.05) are at the noise floor of standard spectrophotometers.

What this calculates

Concentration µg/mL
Analyte concentration in the original sample: A260 × dilution factor × conversion factor.
Copies per µL copies/µL
Template molecules per microlitre of stock (copy-number mode): (ng/µL × 6.022×10²³) / (length × molar mass per base × 10⁹).
Copies per µL (scientific) copies/µL
The copy-number result in scientific notation, with the 10-fold standard series down to 100 copies/µL.
Total yield µg
Total mass of analyte in the final volume (shown when a volume is supplied).
A260/A280 ratio
Purity indicator (shown when A280 is supplied). Typical pure ranges: dsDNA 1.7–2.0, ssDNA 1.8–2.0, RNA 1.9–2.1.

Sources

  • Manchester 1997 Manchester, K.L. (1997). When are nucleic acids not nucleic acids? Problems with estimation of nucleic acid purity by UV absorbance. Biochemical Education 25(4):214–215. — A260/A280 is indicative, not definitive.
  • Sambrook 2001 Sambrook, J. & Russell, D.W. (2001). Molecular Cloning: A Laboratory Manual (3rd ed.). Cold Spring Harbor Laboratory Press. — standard A260 conversion factors (50/33/40 µg/mL per absorbance unit).

Frequently asked questions

  • How do I convert an A260 reading into DNA or RNA concentration?

    Multiply the blank-corrected A260 by the total dilution factor and by the conversion factor for your analyte: concentration (µg/mL) = A260 × dilution factor × factor. The calculator applies the right factor automatically once you pick the analyte, and reports the concentration of the original sample. Enter the final volume as well and it returns the total yield in µg.

  • What is the difference between the 50, 40, and 33 conversion factors?

    These are the standard absorbance constants: a 1.0 A260 reading equals 50 µg/mL for double-stranded DNA, 40 µg/mL for RNA, and 33 µg/mL for single-stranded DNA. The factor you need depends only on the analyte you are measuring, so choose dsDNA, ssDNA, or RNA in the Analyte field. Applying the wrong constant is one of the two classic bench errors this calculator is designed to prevent.

  • What do the A260/A280 and A260/A230 ratios tell me about purity?

    Pure dsDNA reads about 1.8 (range 1.7–2.0), pure RNA about 2.0 (range 1.9–2.1), and ssDNA 1.8–2.0. A ratio at or below 1.6 suggests protein or phenol contamination. The A260/A230 ratio is a second check: pure nucleic acid reads 2.0–2.2, and a lower value points to salt, carbohydrate, or phenol carryover. Supply A280 and the calculator reports the A260/A280 ratio and flags a result outside your analyte range. Treat either ratio as a screen, not a verdict.

  • Do I need to correct for the dilution I made before reading the sample?

    Yes. Enter the total dilution applied to the sample before the reading, and the calculator multiplies it through so the reported concentration refers to your original tube rather than the diluted aliquot. Reading a 2 µL sample into 100 µL, for example, takes a dilution factor of 50. Forgetting that factor is the other classic bench error, and it under-reports concentration by exactly the dilution you applied.

  • What can A260 quantification not tell me about my sample?

    It measures nucleic acid that absorbs near 260 nm, so contaminants such as protein, phenol, or free nucleotides inflate the apparent concentration, and the method cannot tell intact from degraded material. It also assumes a blank-corrected reading taken in a 1 cm path length, or equivalent microvolume optics. Use a gel or fragment-analysis run for integrity, and treat the purity ratio as a coarse screening tool. A low A260/A230 with a normal A260/A280 still points to organic carryover that the 260/280 ratio misses.

For research and educational use only. Not for clinical or diagnostic decisions. Always verify calculations independently before use in critical applications.