Protein Concentration Calculator
Three ways to estimate protein concentration. (1) Beer–Lambert from A280: enter your protein’s molar extinction coefficient and molecular weight to convert A280 into mg/mL. (2) Bradford: paste BSA standards and a sample A595 reading, and the calculator fits the standard curve. (3) BCA: the same standard-curve fit at A562, with a quadratic fit when the response curves. Every fit reports slope, intercept, R² and a plot of the standards.
Local context
Protein concentration is read from absorbance at 280 nm through the aromatic residues (Trp, Tyr, Phe). The A280 method is fast and needs no reagents, which is why it is the first check on a purified sample — and why its limits matter. The Beer-Lambert law gives A = ε × b × c, so c = A / (ε × b). The extinction coefficient ε (M⁻¹cm⁻¹) can be computed from the sequence (Pace et al., 1995) or measured. For 1 mg/mL = 1 AU at 1 cm, ε = 1 / (MW in g/mol) × 1000.
This calculator converts an A280 reading to a concentration using the protein's extinction coefficient (or a sequence-derived estimate from the Edelhoch method) and reports total yield when a sample volume is supplied. It accepts 1 cm or 0.1 cm pathlength, corrects for nucleic acid contamination via A260/A280 if both readings are given, and outputs mg/mL, µg/mL, and µM.
When NOT to use: for crude lysates where Bradford or BCA assay is more reliable (A280 overestimates due to nucleic acids and buffer components). For proteins with no Trp/Tyr (rare), A280 is near zero — use amino acid analysis or quantitative Western. For peptide concentration, the sequence-based ε is less accurate; measure ε experimentally if precision matters.
Key formula: c (mg/mL) = A280 / (ε × b). ε from sequence (Pace et al. 1995) or measured. Reference: Pace et al. (1995) Protein Sci 4:2411; Edelhoch (1967) Biochemistry 6:1948; Thermo Fisher A280 guide.
Calculator
Enter a reading and calculate to see the concentration
Concentration
The sample reading sits outside the standard range, so this result is an extrapolation. Dilute the sample into range and re-measure.
c = A280 × dilution ÷ E(1 mg/mL) · c(M) = A280 × dilution ÷ ε
Beer–Lambert quantification: absorbance at 280 nm is proportional to concentration through the protein's extinction coefficient. In molar mode the result is also converted to mg/mL using the molecular weight.
Related guides
- A280 Protein: Beer-Lambert & ExtinctionHow to calculate protein concentration from A280 absorbance. Covers the Beer-Lambert formula c = A × dilution ÷ E, mass versus molar extinction coefficients, worked examples, and common protein coefficients.
- Calibration Curve: R² & UncertaintyHow to build a calibration curve, assess linearity with R², apply the Beer-Lambert law in quantitative analysis, and propagate measurement uncertainty through dilutions and calculations.
- Spectrophotometry: Beer-Lambert & UV-VisHow spectrophotometry works, the Beer-Lambert law, A260/A280 purity ratios, and how to read absorbance measurements at the bench.
About this calculation
What this is
Proteins absorb at 280 nm because of aromatic residues (tryptophan, tyrosine, and to a lesser extent cystine). For a purified protein with a known molar extinction coefficient, Beer–Lambert gives an accurate concentration; for a mixed or unknown sample, the heuristic 1 A280 = 1 mg/mL is a usable estimate. Bradford and BCA work differently: a dye or copper reaction turns the amount of protein into a colour change, and the sample is read off a curve built from standards of known concentration.
Why it is used
Almost every downstream application (gel loading, dilution to a working stock, assay setup) needs a protein concentration. A280 is the fastest available measurement, but the assumption you make (1 mg/mL per A280 vs. Beer–Lambert) drives the accuracy of every result downstream. Bradford and BCA are the choice when the sample is a mixture, when nucleic acids would inflate an A280 reading, or when the protein has no known extinction coefficient.
Why this calculator exists
A280 readings are often reported in instrument software, but converting them into a usable mg/mL number is where mistakes happen, particularly the missing division by path length or the wrong extinction coefficient. The calculator does the unit bookkeeping and shows both the 1-A280 and Beer–Lambert results side by side. For Bradford and BCA it also fits the standard curve, so the slope, intercept and R² are visible instead of hidden in a spreadsheet.
Worked example
Beer-Lambert at 280 nm is concentration = A280 divided by the extinction coefficient, at 1 cm path length. For BSA the mass coefficient E(1 mg/mL) is 0.66, so an A280 of 0.5 gives 0.5 / 0.66 = 0.76 mg/mL.
The molar route gives the same answer. BSA has a molar extinction coefficient of 43,824 M^-1 cm^-1 and a molecular weight of 66,430, so A280 0.5 is 0.5 / 43,824 = 11.41 µM, and 11.41 µM × 66,430 g/mol = 0.76 mg/mL. If the two routes disagree, one of the coefficients is wrong.
The table lists BSA concentrations for common A280 readings next to the rough rule that 1 A280 equals 1 mg/mL. The rule is close for a mixed sample but it underestimates BSA by about a third, so use the real coefficient for a purified protein.
| A280 (1 cm) | BSA mg/mL (E = 0.66) | Rule of thumb mg/mL |
|---|---|---|
| 0.10 | 0.15 | 0.10 |
| 0.25 | 0.38 | 0.25 |
| 0.50 | 0.76 | 0.50 |
| 1.00 | 1.52 | 1.00 |
| 2.00 | 3.03 | 2.00 |
Common mistakes
- Using an extinction coefficient for a different protein. The coefficient is specific to the protein, the buffer, and the protein form. Using another scales the result in direct proportion.
- Ignoring nucleic acid contamination. Check the A260/A280 ratio. For a clean protein it is near 0.57, and values well above that indicate carry-over. Switch to Bradford or BCA if contamination is suspected.
- Reading a turbid or particulate sample. Centrifuge or filter before reading. Turbid samples scatter light and read artificially high.
- Assuming a 1 cm path length without checking. The formula assumes a 1 cm path length, or an instrument-corrected equivalent, and a blank-corrected reading.
- Using A280 for a very low concentration sample. Use a fluorescence-based assay instead. A280 cannot be used below the instrument noise floor.
- Using the mass coefficient when the molar one is needed. Match the mode to the coefficient you have. The mass coefficient E (1 mg/mL) gives mg/mL, and the molar coefficient ε with the molecular weight gives the molar concentration.
Key assumptions
- The spectrophotometer was blanked with the same buffer the sample was diluted into.
- For the 1-A280 path, the protein mixture is roughly "average" (e.g. a crude lysate); for purified proteins use the calculated path.
- Path length is 1 cm, or the instrument has been corrected for a different path length. For Bradford and BCA, the standards were prepared in the same buffer as the samples and read at the same wavelength (595 nm for Bradford, 562 nm for BCA), and the sample reading falls inside the standard range; readings outside it are flagged as extrapolation.
How to use it
Enter the blank-corrected A280 reading, then choose which coefficient you have. Either a mass extinction coefficient E(1 mg/mL), or a molar extinction coefficient ε together with the molecular weight.
Use the coefficient that matches your protein and its buffer. The coefficient is protein-specific and changes with the reducing state of the sample, so take it from the same source you will quote.
Include the total dilution applied before the reading. If you measured a 1-in-10 dilution, enter a dilution factor of 10.
Read the result in the unit you need. The calculator reports mg/mL and, in molar mode, µM — the two are linked by the molecular weight, so switching between them does not require another measurement.
Blank against the same buffer the protein is in. A buffer with a different absorbance changes the reading, and the blank must cancel that.
If the reading is very high or the buffer is complex, dilute and re-measure, then check the result against a second method. A280 is fast, but a single absorbance number is not a full purity assessment.
For Bradford or BCA, paste the standards first: one line per standard, concentration in mg/mL followed by its absorbance. Then enter the sample absorbance and the dilution factor. Check the R² and the plot before trusting the result; a low R² or a sample outside the standard range means the curve needs to be repeated.
Limitations
- The 1 A280 = 1 mg/mL assumption fails badly for samples rich in nucleic acids (use A260/A280 subtraction).
- It cannot detect proteins that lack aromatic residues or are present at sub-A280 noise-floor concentrations. Bradford responds weakly to some proteins and is disturbed by detergents such as SDS; BCA is disturbed by reducing agents such as DTT and by chelators such as EDTA.
- It does not distinguish folded from denatured protein: the extinction coefficient assumes the folded environment. A standard curve is only as good as its standards: a curve built from two points, or from standards outside the linear range, gives a fragile result.
What this calculates
- Concentration mg/mL
- Protein concentration in the original (undiluted) sample.
- Molar concentration µM
- Molar concentration: shown in ε + molecular-weight mode.
- Standard curve R²
- How well the fitted standard curve explains the standards: shown for Bradford and BCA.
- Standard curve slope
- Fitted slope of absorbance against concentration: shown for a linear Bradford or BCA fit.
- Standard curve intercept
- Fitted absorbance at zero concentration: shown for a linear Bradford or BCA fit.
Sources
- Pace 1995 Pace, C.N., Vajdos, F., Fee, L., Grimsley, G. & Gray, T. (1995). How to measure and predict the molar absorption coefficient of a protein. Protein Science 4(11):2411–2423.
- Sambrook 2001 Sambrook, J. & Russell, D.W. (2001). Molecular Cloning: A Laboratory Manual (3rd ed.). Cold Spring Harbor Laboratory Press. — A280 protein quantitation, Appendix 8.
- Bradford 1976 Bradford, M.M. (1976). A rapid and sensitive method for the quantitation of microgram quantities of protein utilizing the principle of protein-dye binding. Analytical Biochemistry 72:248–254.
- Smith 1985 Smith, P.K. et al. (1985). Measurement of protein using bicinchoninic acid. Analytical Biochemistry 150(1):76–85.
Frequently asked questions
How do I calculate protein concentration from an A280 reading?
The calculator applies the Beer–Lambert law to a blank-corrected A280: divide the absorbance by your extinction coefficient and multiply by the dilution factor. Provide either the mass coefficient E(1 mg/mL) to get mg/mL directly, or the molar coefficient ε together with the molecular weight to get both mg/mL and µM. Enter 1 as the dilution factor for an undiluted reading.
Should I use E(1 mg/mL) or the molar extinction coefficient?
Use whichever coefficient you actually have. E(1 mg/mL) is the absorbance of a 1 mg/mL solution in a 1 cm path, so it yields concentration in mg/mL on its own: BSA, for example, is about 0.66 (E 1 mg/mL at 280 nm); IgG is about 1.4. Use the coefficient for your own protein from the source you will quote. Do not use BSA's value for another protein. The molar coefficient ε in M⁻¹·cm⁻¹ needs the molecular weight in g/mol as well, and it is the option that also reports the concentration in µM.
What does the A280 value actually measure in a protein sample?
Proteins absorb UV light at 280 nm mainly through their tryptophan, tyrosine, and cystine residues, so the absorbance scales with how much protein is in the path. That relationship is only fixed once you supply the coefficient specific to your protein, which is why the calculator asks for E(1 mg/mL) or ε. The upside is that the method needs only a few microlitres and leaves the sample usable for downstream work.
Why does the result show both mg/mL and µM, and when should I use each?
Both appear when you work in the molar mode, because the molecular weight lets the calculator convert between mass and amount. mg/mL is the practical unit for diluting a stock to a working concentration or loading a gel, while µM matters when molar stoichiometry drives the experiment, such as binding or enzyme assays. In E(1 mg/mL) mode only mg/mL is reported, since no molecular weight was entered.
What will make my A280 protein concentration wrong?
Nucleic acid contamination inflates A280 and pushes the calculated concentration high, so check the A260/A280 ratio of the sample and correct if needed. Buffer components that absorb near 280 nm, such as DTT at high concentration, imidazole, or Triton, bias the reading as well, so blank against the same buffer. The calculator assumes a 1 cm path length and a coefficient that matches your protein and its reducing state, and it applies no contamination correction of its own.
For research and educational use only. Not for clinical or diagnostic decisions. Always verify calculations independently before use in critical applications.