The Beer-Lambert law

Absorbance (A) is proportional to concentration (c), path length (l), and the absorbing species: A = ε × c × l, where ε is the molar extinction coefficient. For a given substance at a given wavelength and path length, absorbance changes in direct proportion to concentration.

On the bench, this means you can determine an unknown concentration from an absorbance reading if you know the extinction coefficient and the path length. Rearranged for concentration: c = A ÷ (ε × l).

Absorbance at 260 nm (nucleic acids)

Nucleic acids absorb strongly at 260 nm. For double-stranded DNA, an A260 of 1.0 in a 1 cm cuvette corresponds to about 50 micrograms per mL. For RNA the factor is about 40 micrograms per mL, and for single-stranded DNA about 33 micrograms per mL.

Absorbance at 260 nm also picks up RNA, free nucleotides, and other contaminants, so absorbance methods overestimate DNA in impure samples. Fluorescence methods (Qubit, PicoGreen) are more selective for double-stranded DNA.

One A260 unit (1 cm path) means:

Nucleic acidConcentration at A260 = 1.0 (1 cm path)
Double-stranded DNA50 µg/mL
RNA40 µg/mL
Single-stranded DNA33 µg/mL

Absorbance at 280 nm (protein)

Protein absorbs at 280 nm mainly through the aromatic amino acids tryptophan and tyrosine. An A280 of 1.0 in a 1 cm cuvette corresponds to about 1.0 mg per mL for a typical protein, though the exact value depends on the protein's tryptophan and tyrosine content.

Use the published extinction coefficient (E(1 mg/mL) or ε) for your specific protein. The Protein Concentration calculator converts an A280 reading into concentration in mg per mL or micromolar.

A260/A280 purity ratio

The A260/A280 ratio checks nucleic acid purity. About 1.8 indicates clean double-stranded DNA; about 2.0 indicates clean RNA. A lower value suggests protein or phenol contamination; a higher value can mean RNA contamination.

The ratio is only an indicator. It does not tell you what the contaminant is, and it cannot tell you whether your sample is suitable for a downstream assay.

Read the ratio like this:

A260/A280Likely meaning
~1.8Clean double-stranded DNA
~2.0Clean RNA
Below ~1.6 (DNA sample)Protein or phenol contamination
Above ~2.0 (DNA sample)RNA contamination
~0.57 (protein sample)Clean protein, no nucleic acid

Blanking and path length

Always blank the spectrophotometer with the solvent or diluent your sample is in. This sets the baseline so you measure the sample, not the solvent.

Most cuvettes have a 1 cm path length. Microvolume spectrophotometers use 1 mm or 0.5 mm paths, and the instrument software applies the path length correction automatically. Enter the correct path length for your instrument.

Worked example: DNA concentration from A260

Suppose a double-stranded DNA sample reads A260 = 0.24 in a 1 cm cuvette after blanking. Using the standard conversion of 50 micrograms per mL per absorbance unit, the concentration is 0.24 × 50 = 12 micrograms per mL. If the same sample reads A280 = 0.13, the A260/A280 ratio is 0.24 ÷ 0.13 ≈ 1.85, close to the 1.8 expected for clean DNA.

The calculators apply this arithmetic for you. The Nucleic Acid Quantification calculator takes A260, path length and dilution factor, then returns concentration and the A260/A280 ratio; the Protein Concentration calculator does the same at 280 nm using your protein's extinction coefficient.

Nucleic Acid Quantification

Common mistakes

Do not confuse solvent volume with final solution volume when preparing samples.

Do not skip the mix step; an un-mixed sample gives an uneven reading.

Do not read a sample that contains bubbles or particulates; they scatter light and read artificially high.

Do not compare absorbance readings taken at different path lengths without correcting for path length.

Link to the calculators

The BiomeBench Nucleic Acid Quantification calculator converts A260 absorbance into DNA or RNA concentration in micrograms per mL, applies dilution factors, reports the A260/A280 purity ratio, and calculates total yield.

The Protein Concentration calculator converts an A280 reading into protein concentration using either the mass coefficient (E(1 mg/mL)) or the molar extinction coefficient (ε) and molecular weight.